Aliasing and the Nyquist limit in pulsed-wave Doppler
Two panels show the same Doppler signal sampled at the pulse repetition frequency. On the left the sampling rate is at least twice the maximum Doppler shift, so the samples capture every cycle and the true velocity is displayed. On the right too few samples per cycle are taken, so the reconstructed trace is a false slow wave that wraps across the baseline. This is aliasing, which occurs when the Doppler shift exceeds the Nyquist limit, equal to half the pulse repetition frequency. Below, the fixes are listed: raise the pulse repetition frequency or use high pulse repetition frequency mode, lower the transmit frequency, or increase the insonation angle.