Effective dose from DLP (E = k x DLP)
Two panels. Panel one: the fundamental definition, effective dose E equals the sum over irradiated tissues of tissue weighting factor w-T times equivalent dose H-T; example tissues lung, breast and colon each have w-T about 0.12. Panel two: the practical CT estimate, CTDIvol in mGy times scan length L gives DLP in mGy per cm, then times a conversion coefficient k gives effective dose E in mSv, so k equals E divided by DLP. A bar chart shows k by body region: head about 0.002 (lowest), chest about 0.014, abdomen about 0.015, pelvis about 0.015 to 0.02 (highest). Notes state DLP uses CTDIvol not CTDIw, estimates effective dose not organ absorbed dose, and that k depends on body region and scanner design.