Gamma camera QC: flood-field uniformity & artefacts

Two circular flood-field images from a gamma camera. The left field shows an even, uniform density of counts and is marked acceptable. The right field shows the same density interrupted by a focal cold spot caused by a cracked crystal or photomultiplier drift. Below is a list of gamma-camera quality-control tests: uniformity uses a flood-field acquisition with a uniform planar source and should be a few percent, not twenty percent; spatial resolution uses a bar phantom or point or line source measured as full width at half maximum; linearity uses a line source; and intrinsic tests are done without a collimator while extrinsic tests include the collimator, so intrinsic resolution is better than system resolution.

Gamma camera flood-field uniformity: good vs non-uniform, with the QC test list Two circular flood-field images from a gamma camera. The left field shows an even, uniform density of counts and is marked acceptable. The right field shows the same density interrupted by a focal cold spot caused by a cracked crystal or photomultiplier drift. Below is a list of gamma-camera quality-control tests: uniformity uses a flood-field acquisition with a uniform planar source and should be a few percent, not twenty percent; spatial resolution uses a bar phantom or point or line source measured as full width at half maximum; linearity uses a line source; and intrinsic tests are done without a collimator while extrinsic tests include the collimator, so intrinsic resolution is better than system resolution. Flood-field uniformity: uniform planar source imaged over the field of view Good: even count density integral uniformity a few percent Non-uniform: focal cold spot cracked crystal or PMT drift What each QC test uses Uniformity flood-field, uniform planar source; a few percent, not 20 percent Spatial resolution bar phantom or point / line source; measured as FWHM Linearity, distortion line source: width and straightness of the imaged line Intrinsic vs extrinsic no collimator (Co-57 sheet) vs with collimator (Tc-99m flood) Resolution rule intrinsic beats system; Rsys = sqrt(Rint squared + Rcoll squared)

Practice the Physics paper

FRCRX is a Part 1 Physics question bank: true/false items with explanations, timed mocks and topic analytics.