Mammography vs conventional radiography technique factors
Two-column comparison. Mammography uses low tube voltage (about 25 to 32 kVp) giving a lower mean photon energy (about 17 to 23 keV) with molybdenum or rhodium and tungsten target and filter combinations, a small focal spot (about 0.1 to 0.3 mm), a short focus-to-image distance (about 60 to 65 cm), longer exposure times, an anti-scatter grid in contact views, and contrast driven by the photoelectric effect. General radiography uses 60 to 120 kVp, a harder beam, a tungsten target, a larger focal spot, a longer focus-to-image distance (about 100 to 115 cm), shorter exposures, and relatively more Compton scatter. The goal in mammography is high soft-tissue contrast and high spatial resolution to resolve microcalcifications.