PET coincidence detection & annihilation geometry
A positron from beta-plus decay travels a short range in tissue and annihilates with an electron, producing two 511 keV photons emitted about 180 degrees apart. A circumferential ring of scintillation detectors registers the two photons within a short coincidence timing window; the pair of struck detectors defines a line of response. A lower panel contrasts true coincidences (both photons from one event, line of response passes through it), scatter coincidences (one photon is Compton-scattered so the line of response is mispositioned), and random coincidences (single photons from two separate events fall in the window, giving a false line of response).