Infographics / Ultrasound Physics
Doppler equation and insonation angle
An ultrasound beam from a transducer insonates a blood vessel at angle theta to the flow direction. The detected Doppler shift depends only on the along-beam velocity component v times cosine theta, following the equation delta f equals 2 f0 v cosine theta divided by c. A graph shows the shift is maximum at 0 degrees (beam parallel to flow), falls to zero at 90 degrees (perpendicular), and is most accurate below 60 degrees.