Patient dose vs mAs, kV and pitch
A graph of relative patient dose against relative mAs (tube current times rotation time). The dose against mAs relationship is a straight line through the origin: doubling mAs doubles the dose. For contrast, a second dashed curve shows tube output rising steeply with kVp (about kVp squared to cubed), which is non-linear. A side note states noise falls as one over the square root of mAs, so doubling mAs reduces noise only to about 0.71, not half.